Product Name:
Mouse C-X-C Motif Chemokine 5 (CXCL5) ELISA Kit
Cat.#:
JL28925
Regulatory Status:
Alternative:
CXCL5; ENA78; SCYB5
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
62.5-4000pg/mL
Sensitivity:
29.5pg/mL
Specificity:
This kit recognizes Mouse CXCL5 in samples. No significant cross-reactivity or interference between Mouse CXCL5 and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Research Area:
Immunology
Product Datasheets
| CXCL5
C-X-C motif chemokine 5 (CXCL5) is a small cytokine belonging to the CXC chemokine family that is also known as epithelial-derived neutrophil-activating peptide 78 (ENA-78). It is produced following stimulation of cells with the inflammatory cytokines interleukin-1 or tumor necrosis factor-alpha.Expression of CXCL5 has also been observed in eosinophils, and can be inhibited with the type II interferon IFN-γ.This chemokine stimulates the chemotaxis of neutrophils possessing angiogenic properties. It elicits these effects by interacting with the cell surface chemokine receptor CXCR2.The gene for CXCL5 has four exons and is located on human chromosome 4 amongst several other CXC chemokine genes.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 4000 | 2000 | 1000 | 500 | 250 | 125 | 62.5 | 0 |
| OD | 2.22 | 1.56 | 0.92 | 0.63 | 0.42 | 0.25 | 0.18 | 0.08 |
| Corrected OD | 2.14 | 1.48 | 0.84 | 0.55 | 0.34 | 0.17 | 0.1 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Mouse CXCL5 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-101 | 96 |
| Plasma(n=8) | 92-105 | 102 |
| Cell Culture Supernates(n=8) | 96-108 | 105 |
| Linearity
Samples were spiked with high concentrations of Mouse CXCL5 and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-96 | 90-110 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-103 | 87-108 | 105-115 |
| Average (%) | 94 | 98 | 108 |
Size
Quantity
Price ($)
48T
1
360 / 360
96T
1
640 / 640