Product Name:
Human Coagulation Factor Ⅹ (F10) ELISA Kit
Cat.#:
JL12112
Regulatory Status:
Alternative:
FX; FXA; Stuart Prower Factor; Thrombokinase
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.94-60ng/mL
Sensitivity:
0.48ng/mL
Specificity:
This kit recognizes Human F10 in samples. No significant cross-reactivity or interference between Human F10 and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Research Area:
Cardiovascular Biology
Product Datasheets
| F10
Coagulation factor X (EC 3.4.21.6), or Stuart factor, is an enzyme of the coagulation cascade.It is a serine endopeptidase (protease group S1, PA clan). Factor X is synthesized in the liver and requires vitamin K for its synthesis.Factor X is activated, by hydrolysis, into factor Xa by both factor IX with its cofactor, factor VIII in a complex known as intrinsic pathway; and factor VII with its cofactor, tissue factor in a complex known as extrinsic pathway.It is therefore the first member of the final common pathway or thrombin pathway.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 60 | 30 | 15 | 7.5 | 3.75 | 1.87 | 0.94 | 0 |
| OD | 2.22 | 1.56 | 0.92 | 0.63 | 0.42 | 0.25 | 0.18 | 0.08 |
| Corrected OD | 2.14 | 1.48 | 0.84 | 0.55 | 0.34 | 0.17 | 0.1 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human F10 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-102 | 95 |
| Plasma(n=8) | 92-106 | 103 |
| Linearity
Samples were spiked with high concentrations of Human F10 and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-95 |
| Average (%) | 91 | 93 | |
| 1 : 4 | Range (%) | 89-103 | 87-108 |
| Average (%) | 94 | 97 |
Size
Quantity
Price ($)
48T
1
280 / 280
96T
1
460 / 460