400-0066-400

Human Cluster Of Differentiation 200 (CD200) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Cluster Of Differentiation 200 (CD200) ELISA Kit
Product Name:

Human Cluster Of Differentiation 200 (CD200) ELISA Kit

Cat.#:

JL29492

Regulatory Status:
Alternative:

MOX1; MOX2; MRC; OX-2; My033; Antigen Identified By Monoclonal Antibody MRC OX-2

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
31.2-2000pg/mL
Sensitivity:
15.2pg/mL
Specificity:
This kit recognizes Human CD200 in samples. No significant cross-reactivity or interference between Human CD200 and analogues was observed.
Sample Type:
Serum;plasma and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Immunology
  • Human Cluster Of Differentiation 200 (CD200) ELISA Kit
  • Human Cluster Of Differentiation 200 (CD200) ELISA Kit
  • Human Cluster Of Differentiation 200 (CD200) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| CD200

OX-2 membrane glycoprotein, also named CD200 (Cluster of Differentiation 200) is a protein encoded by the CD200 gene. The CD200 gene is located on chromosome 3 in proximity to genes encoding the other B7 proteins CD80/CD86. In mice, the CD200 gene is located on chromosome 16.The protein encoded by this gene is a type-1 membrane glycoprotein, which contains two IgSF immunoglobulin domains, transmembrane region and a 19 amino acid long cytoplasmatic domain. CD 200 belongs to the immunoglobulin superfamily, particularly belongs to the B7 receptor family.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 2000 1000 500 250 125 62.5 31.2 0
OD 2.28 1.91 1.43 0.97 0.62 0.38 0.22 0.07
Corrected OD 2.21 1.84 1.36 0.9 0.55 0.31 0.15 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human CD200 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102





 

| Linearity

Samples were spiked with high concentrations of Human CD200 and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4)
1 : 2 Range (%) 84-95 88-96
Average (%) 91 92
1 : 4 Range (%) 89-103 87-108
Average (%) 94 98







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
360 / 360
96T
1
640 / 640
Top