Product Name:
Human Hematopoietic Prostaglandin D Synthase (PGDS) ELISA Kit
Cat.#:
JL21967
Regulatory Status:
Alternative:
HPGDS; GSTS; H-PGDS; PDS; PGD2; PGDS; PGDS2; Glutathione S-Transferase Sigma; Glutathione-requiring prostaglandin D synthase; Prostaglandin D2 Synthase; Hematopoietic
Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
78.1-5000pg/mL
Sensitivity:
40pg/mL
Specificity:
This kit recognizes Human PGDS in samples. No significant cross-reactivity or interference between Human PGDS and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Research Area:
Metabolism
Product Datasheets
| PGDS
PGDS protein is a protein that is encoded by the HPGDS gene. Prostaglandin-D synthase is a sigma class glutathione-S-transferase family member. The enzyme catalyzes the conversion of PGH2 to PGD2 and plays a role in the production of prostanoids in the immune system and mast cells. The presence of this enzyme can be used to identify the differentiation stage of human megakaryocytes.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 5000 | 2500 | 1250 | 625 | 312.5 | 156.2 | 78.1 | 0 |
| OD | 2.41 | 1.84 | 1.21 | 0.85 | 0.52 | 0.33 | 0.21 | 0.08 |
| Corrected OD | 2.33 | 1.76 | 1.13 | 0.77 | 0.44 | 0.25 | 0.13 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Human PGDS spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 84-101 | 96 |
| Plasma(n=8) | 92-105 | 102 |
| Linearity
Samples were spiked with high concentrations of Human PGDS and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | ||
| 1 : 2 | Range (%) | 84-95 | 88-96 |
| Average (%) | 91 | 93 | |
| 1 : 4 | Range (%) | 89-103 | 87-108 |
| Average (%) | 94 | 98 |
Size
Quantity
Price ($)
48T
1
280 / 280
96T
1
460 / 460