400-0066-400

Human Lactate Dehydrogenase A (LDH-A) ELISA Kit

Home  >  Sandwich ELISA Kits  >  Human Lactate Dehydrogenase A (LDH-A) ELISA Kit
Product Name:

Human Lactate Dehydrogenase A (LDH-A) ELISA Kit

Cat.#:

JL19761

Regulatory Status:
Alternative:

LDH-A; LDH-M; LDH1; PIG19; Cell proliferation-inducing gene 19 protein; LDH muscle subunit; Renal carcinoma antigen NY-REN-59

Detection Method:
ELISA
Assay Type:
Sandwich
Detection Range:
0.15-10ng/mL
Sensitivity:
0.02ng/mL
Specificity:
This kit recognizes Human LDH-A in samples. No significant cross-reactivity or interference between Human LDH-A and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Metabolism
  • Human Lactate Dehydrogenase A (LDH-A) ELISA Kit
  • Human Lactate Dehydrogenase A (LDH-A) ELISA Kit
  • Human Lactate Dehydrogenase A (LDH-A) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| LDH-A

Lactate dehydrogenase A (LDHA) is an enzyme encoded by the LDHA gene. It is a monomer of Lactate dehydrogenase, which exists as a tetramer. The other main subunit is lactate dehydrogenase B (LDHB). LDHA is found in most somatic tissues, though predominantly in muscle tissue and tumors, and belongs to the lactate dehydrogenase family. It has also been shown that LDHA plays an important role in the development, invasion and metastasis of malignancies. Mutations in LDHA have been linked to exertional myoglobinuria.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (ng/mL) 10 5 2.5 1.25 0.62 0.31 0.15 0
OD 2.4 1.77 1.11 0.7 0.45 0.3 0.2 0.07
Corrected OD 2.33 1.7 1.04 0.63 0.38 0.23 0.13 -
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of Human LDH-A spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 84-101 96
Plasma(n=8) 92-105 102
Tissue Homogenates(n=8) 96-108 104





 

| Linearity

Samples were spiked with high concentrations of Human LDH-A and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Tissue Homogenates(n=4)
1 : 2 Range (%) 84-95 88-96 92-110
Average (%) 91 93 96
1 : 4 Range (%) 89-103 87-108 105-115
Average (%) 94 100 108







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
280 / 280
96T
1
460 / 460
Top