Product Name:
Mouse Hepatitis B virus surface antibody (HBsAb) ELISA Kit
Cat.#:
JL20665
Regulatory Status:
Alternative:
Hepatitis B virus surface antibody
Detection Method:
ELISA
Assay Type:
Indirect
Detection Range:
3.12-200ng/mL
Sensitivity:
1.41ng/mL
Specificity:
This kit recognizes Mouse HBsAb in samples. No significant cross-reactivity or interference between Mouse HBsAb and analogues was observed.
Sample Type:
Serum, plasma and other biological fluids
Assay Procedure:
Research Area:
Immunology
Product Datasheets
| HBsAb
Hepatitis B surface antibody is a protective antibody, primarily produced by the body's immune system in response to stimulation by the hepatitis B surface antigen. Its presence can neutralize the hepatitis B virus within the body. When the hepatitis B virus invades the body, it stimulates the immune system to generate an immune response. B lymphocytes in the immune system differentiate into effector B cells (plasma cells), which secrete a specific immunoglobulin G—this is the surface antibody. It can specifically bind to the surface antigen, and then, through combined action with other immune functions in the body, eliminate the virus, rendering the body no longer susceptible to hepatitis B virus infection. Therefore, the surface antibody is referred to as a protective antibody.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (ng/mL) | 200 | 100 | 50 | 25 | 12.5 | 6.25 | 3.12 | 0 |
| OD | 2.52 | 1.96 | 1.38 | 0.97 | 0.55 | 0.31 | 0.19 | 0.08 |
| Corrected OD | 2.44 | 1.88 | 1.3 | 0.89 | 0.47 | 0.23 | 0.11 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Mouse HBsAb spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 88-102 | 97 |
| Plasma(n=8) | 94-107 | 104 |
| Linearity
Samples were spiked with high concentrations of Mouse HBsAb and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | ||
| 1 : 2 | Range (%) | 82-93 | 88-96 |
| Average (%) | 91 | 93 | |
| 1 : 4 | Range (%) | 89-103 | 87-110 |
| Average (%) | 94 | 98 |
Size
Quantity
Price ($)
48T
1
360 / 360
96T
1
640 / 640