Product Name:
Rat Beta-Crosslaps (β-CTx) ELISA Kit
Cat.#:
JL51641
Regulatory Status:
Alternative:
b-CTx; bCTXI; bCTX-I; B-Cr; BCL; Type I Collagen C-Telopeptide-Related Fraction
Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
125-8000pg/mL
Sensitivity:
62.6pg/mL
Specificity:
This kit recognizes Rat β-CTx in samples. No significant cross-reactivity or interference between Rat β-CTx and analogues was observed.
Sample Type:
Serum;plasma;tissue homogenates;cell lysates;cell culture supernates and other biological fluids
Assay Procedure:
Research Area:
Immunology
Product Datasheets
| β-CTx
Beta-CrossLaps (β-CTx), the C-terminal telopeptide of type I collagen, is a specific biomarker released into circulation during bone resorption. As a degradation product of bone collagen matrix, elevated β-CTx levels directly reflect increased osteoclastic activity and bone loss, serving as a sensitive indicator for osteoporosis and Paget's disease. Clinically, β-CTx measurement is primarily used to monitor the efficacy of antiresorptive therapies in metabolic bone disorders. Currently recognized as the gold standard for bone resorption assessment, this marker demonstrates superior specificity compared to other available tests. The serum CrossLaps assay specifically detects these carboxy-terminal collagen crosslinks, which are enzymatically cleaved C-telopeptides from fibrillar collagen during bone remodeling.| Typical data
As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.| Concentration (pg/mL) | 8000 | 4000 | 2000 | 1000 | 500 | 250 | 125 | 0 |
| OD | 2.11 | 1.26 | 0.8 | 0.5 | 0.33 | 0.24 | 0.19 | 0.08 |
| Corrected OD | 2.08 | 1.18 | 0.72 | 0.42 | 0.25 | 0.16 | 0.11 | - |
| Repeatability
Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%| Recovery
The recovery of Rat β-CTx spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.| Sample Type | Range(%) | Average Recovery(%) |
| Serum(n=8) | 88-102 | 97 |
| Plasma(n=8) | 94-107 | 104 |
| Cell Culture Supernates(n=8) | 96-111 | 107 |
| Linearity
Samples were spiked with high concentrations of Rat β-CTx and diluted with Universal Diluent to produce samples with values within the range of the assay.| Serum(n=4) | Plasma(n=4) | Cell Culture Supernates(n=4) | ||
| 1 : 2 | Range (%) | 82-93 | 88-96 | 90-110 |
| Average (%) | 91 | 93 | 96 | |
| 1 : 4 | Range (%) | 89-103 | 87-110 | 105-115 |
| Average (%) | 94 | 98 | 109 |
Size
Quantity
Price ($)
48T
1
280 / 280
96T
1
460 / 460