400-0066-400

6-keto-prostaglandin F1α (6-keto-PGF1α) ELISA Kit

Home  >  Competitive ELISA Kits  >  6-keto-prostaglandin F1α (6-keto-PGF1α) ELISA Kit
Product Name:

6-keto-prostaglandin F1α (6-keto-PGF1α) ELISA Kit

Cat.#:

JL58304

Regulatory Status:
Alternative:

6-keto-PGF1α

Detection Method:
ELISA
Assay Type:
Competitive
Detection Range:
3.12-200pg/mL
Sensitivity:
1.76pg/mL
Specificity:
This kit recognizes 6-keto-PGF1α in samples. No significant cross-reactivity or interference between 6-keto-PGF1α and analogues was observed.
Sample Type:
Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Procedure:
Assay Procedure
Research Area:
Cardiovascular Biology
  • 6-keto-prostaglandin F1α (6-keto-PGF1α) ELISA Kit
  • 6-keto-prostaglandin F1α (6-keto-PGF1α) ELISA Kit
  • 6-keto-prostaglandin F1α (6-keto-PGF1α) ELISA Kit
Size & Price: cart
×
Human A1BG Polyclonal Antibody
Email *
Message *
Please Sign In.
Product Datasheets
Instruction: Down Instruction


| 6-keto-PGF1α

6-keto-prostaglandin F1α (6-keto-PGF1α) is the inactive, non-enzymatic hydrolysis product of PGI2. It serves as a useful biomarker for in vivo PGI2 biosynthesis. When [3H]-PGI2 is injected into healthy males, 6.6% of the radioactivity is recovered in urine as [3H]-6-keto-PGF1α. At physiological pH, prostacyclin rapidly forms the inactive hydrate product 6-keto-prostaglandin F1α‌.
 

| Typical data

As the OD values of the standard curve may vary according to the conditions of the actual assay performance (e.g. operator, pipetting technique, wash technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data are provided below for reference only.

 
Concentration (pg/mL) 200 100 50 25 12.5 6.25 3.12 0
OD 0.13 0.19 0.32 0.56 0.8 1.18 1.47 2.04
 





| Repeatability

Intra-assay Precision (Precision within an assay):CV% < 10%; Inter-assay Precision (Precision between assays):CV% < 10%
 

| Recovery

The recovery of 6-keto-PGF1α spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

 
Sample Type Range(%) Average Recovery(%)
Serum(n=8) 87-98 93
Plasma(n=8) 91-113 101
Cell Culture Supernates(n=8) 81-114 100





 

| Linearity

Samples were spiked with high concentrations of 6-keto-PGF1α and diluted with Universal Diluent to produce samples with values within the range of the assay.

 
    Serum(n=4) Plasma(n=4) Cell Culture Supernates(n=4)
1 : 2 Range (%) 83-111 83-94 82-100
Average (%) 97 86 90
1 : 4 Range (%) 86-113 83-102 83-108
Average (%) 103 91 100







 
Close
Sample Data
Close
Sample Data
Close
Sample Data
Size Quantity Price ($)
48T
1
520 / 520
96T
1
760 / 760
Top